Double exposure

A double exposure is two radiographic images recorded on one receptor, one over the other. UK dental guidance lists it as an operator error under both film and phosphor plate acquisition, and records its effect: superimposed images. Alongside that it records unnecessary dose delivered to the patient.
Synonyms: double exposed film; superimposed exposure; twice-exposed receptor
On this page6 sections
What does a double exposure look like on the image?
Superimposed images. That is the whole of the recorded description, and it is worth being honest that the literature does not go further.
The UK guidance notes on the safe use of X-ray equipment carry an appendix of common image quality faults, laid out as cause and effect. Under exposure faults, for film and for phosphor plates alike, the cause reads "double exposure" and the effect reads "superimposed images". Nothing retrieved for this entry describes the appearance in more detail than that — not the classification studies, not the guidance.
So you are looking for two sets of structures in one picture where there should be one. What that resolves to on any given receptor depends on where the receptor sat for each exposure, and no source puts a shape on it.
One detail does travel with it, and it is the useful one. Guidance lists a separate operator error, "image receptor not exposed", whose recorded effect is no image captured — and it notes that this may occur in tandem with double exposure of another image receptor. If a full-mouth series comes back with one blank receptor, the doubled one may be in the same set.
How does it happen, and which receptors can it happen to?
The receptor is exposed, and then it is exposed again before the first image has been read off it.
Guidance names double exposure in exactly two places. It appears under film-based image acquisition, and it appears under digital image acquisition tagged specifically to CR — computed radiography, the phosphor plate route. It does not appear in the intraoral radiography table at all, which covers preparation, positioning and patient motion. This is filed under image acquisition and processing, not under intraoral radiography.
The solid-state sensor is the interesting absence. In the same digital table, the category for DR image receptors carries two rows and only two: a cracked sensor, and a receptor used back to front, which shows the read-out circuitry of the sensor on the image. There is no double-exposure row under DR sensors, and the double-exposure row that does exist is the one tagged CR.
Be careful about what that does and does not establish. It says that professional guidance does not recognise double exposure as a fault of a rigid sensor. It is not an explanation of the read-out behaviour that makes the difference, and no source retrieved for this entry supplies one. Take it as what the fault taxonomy records, not as physics.
Double exposure is also still a live category in the current phosphor plate literature. A 2026 multi-centre study of 1,497 intraoral phosphor plate radiographs annotated them against a standardised taxonomy of four artifact groups and 29 artifact classes, one of which is named for double exposure and sits in the exposure-related group.
Is an image left on an unerased phosphor plate the same thing?
No — it has its own name, in both channels, and it is classed as a different kind of error.
Guidance calls it partial image clearing and files it as a CR reader error, under image processing rather than under exposure. Its recorded effect is the current and the previously captured image superimposed. A 2017 Turkish classification study, which sorted 34 intraoral phosphor plate errors and artefacts into four causative groups, calls it an erasure artefact and places it among the scanner-induced artefacts, caused by improper erasure and reuse of the plate for another radiograph.
Here is the part worth carrying away. That same study defines the erasure artefact as a double image "which has the same characteristics as an image acquired as a result of a double exposure". Separately named, separately classified, and — on the evidence of the one study that says so directly — not separable by looking. What separates them is the cause, not the picture.
The naming is not perfectly tidy even inside one paper. The 2017 study also lists improper erasure of a previously acquired latent image as one of the causes of a double image among its operator- and patient-induced errors, so the unerased plate turns up twice: once as a cause of a double image, once as its own class. Treat the distinction as real but the boundary as slightly soft.
Both are rare in the counted series. Across 2,100 intraoral radiographs in that 2017 study, the erasure artefact was recorded zero times. In the 2026 multi-centre set, double exposure was among the classes the segmentation model scored zero on, and the authors note that several artifact categories were absent from the training data altogether. That is a statement about how sparse these artifacts are in real archives, not a measured rate.
Guidance is short and specific here: phosphor plates should be erased before they are exposed for uniformity or image quality checks, and before they return to clinical use.
How do you tell it from a ghost image or from superimposed anatomy?
| Looks like | What tells them apart |
|---|---|
| Ghost image | Panoramic only. A rotational panoramic radiograph is a tomographic image plus redundant shadows, and a ghost is a redundant shadow cast when the object lies between the centre of rotation and the source of the beam. An intraoral receptor never has that geometry. |
| Superimposed anatomy | One exposure, not two. It is a projection error from beam angulation — a wrong horizontal beam angle gives overlapping contact points — and it is independent of receptor type. |
| Partial image clearing | Not separable by appearance; the 2017 classification gives it the same characteristics as a double exposure. It is separated by cause: a reader error on a plate that was not fully cleared. |
Two cautions on wording. In panoramic radiography "double image" is a term of art for a third, distinct thing — the panoramic literature describes real, ghost and double images as three separate sources of shadow — so it is not a synonym for double exposure. And the panoramic ghost has its own guidance entry under positioning, where a patient not standing straight in the machine produces an exaggerated ghost shadow of the cervical spine over the anterior teeth.
Can it be read, or does it always need a retake?
Nothing retrieved for this entry says a double exposure always needs a retake, and it is worth resisting the temptation to say so.
What guidance gives instead is a criterion that applies to every image. Dental radiographs and CBCT images are rated on a two-point scale, diagnostically acceptable or not acceptable. Acceptable means no errors or minimal errors in patient preparation, exposure, positioning, image receptor processing or image reconstruction, and of sufficient image quality to answer the clinical question. Not acceptable means errors in those same things which render the image diagnostically unacceptable. The recommended targets are at least 95% acceptable for digital imaging and at least 90% for film.
So the question is not whether the artifact is present. It is whether the clinical question can still be answered.
What guidance asks for, whenever an image is rated not acceptable, is a record: the date, the nature of the deficiency, its known or suspected cause, and the number of repeat radiographs, if any were taken. The repeat is written as conditional. The record is not.
For scale, retakes for image artifacts are ordinary rather than exceptional. In a two-year series of 50,000 intraoral digital radiographs from a single university clinic in Saudi Arabia, 3,550 — 7.1% — were retaken because of image artifacts, and periapical radiographs accounted for 80.8% of those retakes.
The wider point sits with the other faults in the radiographic artifact hub, alongside the herringbone artifact and the wider problem of telling an artifact from real pathology: most of them you learn to read through. This one you learn to recognise, because recognising it is what tells you which exposure to repeat and which record to write.
Sources and further reading
6 sources
- Çalışkan A, Sumer AP. Definition, classification and retrospective analysis of photostimulable phosphor image artefacts and errors in intraoral dental radiography. Dentomaxillofacial Radiology. 2017;46(3):20160188. DOI
- Guidance Notes for Dental Practitioners on the Safe Use of X-ray Equipment, 2nd edition. Public Health England, with the Faculty of General Dental Practice (UK), Royal College of Surgeons of England; 2020. Source
- Topal CG, Yalçın O, Tetik H, Ünal M, Bandirmali Erturk N, Üçok CÖ. Automated segmentation of digital artifacts in intraoral photostimulable phosphor radiographs. Diagnostics. 2026;16(8):1194. DOI
- Elkhateeb SM, Aloyouny AY, Omer MMS, Mansour SM. Analysis of photostimulable phosphor image plate artifacts and their prevalence. World Journal of Clinical Cases. 2022;10(2):437–447. DOI
- Wakoh M, Kuroyanagi K. Mathematical approach to horizontal and vertical magnification factors in rotational panoramic radiography — with attention to redundant shadows. Bulletin of Tokyo Dental College. 1991;32(3):87–94. PubMed
- Wakoh M, Kuroyanagi K. Redundant shadows in rotational panoramic radiographs. Estimating the imaging positions of objects by mathematical analysis. Oral Surgery, Oral Medicine, Oral Pathology. 1991;72(4):497–504. DOI